HVC802A0 X-Ray Source

Key Features:
●Integrated design with high electrical integration and a compact appearance
●Capable of continuous, uninterrupted operation for extended periods
●High stability
●Can be installed in any orientation
●Standard digital interface RS232, easy to use
  • Detail
  • Parameters

Product Details

The HVC802A0 X-Ray Source is an exceptionally compact, portable, and reliable line of high voltage X-ray generators. Using in allowable operating ambient temperature, it is capable to self-heat dissipate up to 100W continuous with no additional cooling. The X-ray generator integrates the high voltage inverter, filament supply, X-ray tube and X-ray housing into one unit. The High Voltage X-ray Unit is an 80kV/100W, High Frequency, self-cooled, self-protected & self-contained X-ray generator. It consists of an X-ray tank, and a control box. The Generator is controlled, programmed and monitored via a standard RS232 interface.

Application:

Food testing, industrial non-destructive testing, dangerous goods testing and other fields, mostly used for simple X-ray machines or mobile X-ray testing equipment

Specification:

Item

Specification

Input voltage

230VAC±10%, 50/60Hz, 1.2Amps

Output power of X ray tube

Max continuous output power 100W80kV/1.25mA,40kv/2.5mA

Output voltage

Rated output voltageContinuously adjustable voltage range 20kV—80kV

Output voltage ripple±0.5%peak to peak

Output voltage accuracy±1% of voltage setting value

line regulation±0.1%

load regulation±0.1%

Tube current

Rated tube currentContinuously adjustable current range 0.2mA-2.5mA

Tube current accuracy±1% mA of current setting value

line regulation±0.5%

load regulation±0.5%

Rise time of output voltage

The kV rise time is <0.4 Sec from 10% to 90% of the output voltage.

Filament power supply:

input voltage24VDC

filament voltage2.0 to 3Vac

Filament current3.0 to 3.5 Amps RMS

preheating time3sec

Tube feature

Tube typefixed anode, glass window, tungsten target

focus0.8mm

inherent filtration0.8mm Be0.7mm AI

radiation angle80°*16° fan beam

target angle25°

Cooling

transformer oilnatural air cooling

Working temperatures

-10℃—40℃

Storing temperature

-20℃—60℃

System temperature protection

60°C ± 3°C of Oil temperature:

Humidness

98%Non-condensation

Weight

23kg

Installation direction

Installation in any direction

Radiation angle

16°×80°

X-ray leakage

Less than 0.5mR/hr at 5cm from the surface of the HVC802A0.

JB1/AC~(AC Input Power Connector)

Pin

Signal

Parameter

1

L

live wire

2

N

Neutral line

3

G

PE

1.jpg

JB2/COM (DMR-9S interface definitions)

Pin

Signal

Parameter

1.4.6.7.8.9

N/C

No connect

2

TXD

Data transmit

3

RXD

Data receive

5

GND

Signal gnd

2.png

JB3/Interlock,(DMR-9P interface definitions)

Pin

Signal

Parameter

3/4/5/6/7/8/9

N/C

No connect

1

Interlock Out


2

Interlock In


3.png

Short connect pin1 and pin2 make X ray source normal operation. Typical connection

4.png

Led indicator

ID

Color

Meaning

XrayOn

Yellow

indicate X ray on

ARC

Red

Arcing in oil tank

OT

Red

Over temperature

EP_Err

Red

Tube voltage error

IP_Err

Red

Tube current error

Power

Green

Power on

Tank size

5.png


Home