HVC1205A0 X-Ray Source

Key Features:
●Compact design, easy installation
●System integrates high-voltage power supply, filament power supply, and control unit
●Self-circulating cooling system
●Versatile installation options
●Standard digital interface for simple application
  • Detail
  • Parameters

Product Details

The HVC1205A0 is a compact, safe, low-leakage, and electrically stable integrated X-ray source. It can operate continuously at 500W within the allowable temperature range. The maximum output voltage can reach 120kV/500W. The integrated X-ray source mainly consists of a control box and high-voltage oil tank. The system uses RS232 interface for control, monitoring, and firmware upgrades.

Application:

Food testing, industrial non-destructive, security application and other fields.

Specification:

Item

Specification

Input voltage

230VAC±10%, 50/60Hz, 4.2Amps

Output power of X ray tube

Max continuous output power 500W100kV/5mA or 120kV/4.1mA

Output voltage

Rated output voltageContinuously adjustable voltage range 30kV—120kV

Output voltage ripple±0.5%peak to peak

Output voltage accuracy±2% of voltage setting value

line regulation±0.1%

load regulation±0.1%

Tube current

Rated tube currentContinuously adjustable current range 0.2mA-5.0mA

Tube current accuracy±0.2mA of current setting value

line regulation±0.1%

load regulation±0.1%

Rise time of output voltage

The kV rise time is <0.6 Sec at maximum power

The kV rise time is <0.1 Sec at low voltage ( < 40kV )

Filament power supply:

input voltage24VDC

filament voltage3.7Vac

filament3.0 to 3.5 Amps RMS

preheating time3sec

Tube feature

Tube typefixed anodeglass envelopetungsten target

focus0.8mm

inherent filtration0.8mm Be0.7mm AI

radiation angle80°×36°fan beam

target angle25°

Cooling

transformer oilexternal circulationexternal forced air cooling

Working temperatures

-10℃—40℃

Storing temperature

-20℃—60℃

System temperature protection

60±3℃ of oil temperature

Humidness

98%Non-condensation

Weight

Oil tank28.5kg

Control Unit3.8kg

Installation direction

Installation in any direction

Radiation angle

36°×80°

X-ray leakage

Less than 0.5mR/hr @5cm from the surface of the HVC1205A0.

JB1/AC~(AC Input Power Connector)

Pin

Signal

Parameter

1

L

live wire

2

N

Neutral line

3

G

PE

1.jpg

JB2/COM (DMR-9S interface definitions)

Pin

Signal

Parameter

1.4.6.7.8.9

N/C

No connect

2

TXD

Data transmit

3

RXD

Data receive

5

GND

Signal gnd

2.png

JB3/Interlock,(DMR-9P interface definitions)

Pin

Signal

Parameter

3/4/5/6/7/8/9

N/C

No connect

1

Interlock Out


2

Interlock In


3.pngShort connect pin1 and pin2 make X ray source normal operation. Typical connection

4.png

Led indicator:

ID

Color

Meaning

XrayOn

Yellow

indicate X ray on

ARC

Red

Arcing in oil tank

OT

Red

Over temperature

EP_Err

Red

Tube voltage error

IP_Err

Red

Tube current error

Power

Green

Power on

5.png

Home