HVC80804 X-Ray Source

Key Features:
●Integrated design with a compact structure and easy installation
●Equipped with an external beryllium window tube, ideal for low-voltage applications
●High system integration with a self-circulating cooling system
●Versatile installation options
●Standard digital interface for simple application
  • Detail
  • Parameters

Product Details

The HVC80804 is a compact, safe, low-leakage, and electrically stable integrated X-ray source. It can operate continuously at 350W within the allowable temperature range. The system includes a high-voltage power supply, filament power supply, X-ray tube, high-voltage oil tank, and radiator. This high-frequency, self-cooling, and self-protecting X-ray generator delivers up to 80kV and 350W. It primarily consists of a control box and high-voltage oil tank, with RS232 interface for system control, monitoring, and firmware upgrades.

Application:

Food testing, Shoes and clothing detection, fluorescence analysis application and other fields, mostly used for X-ray machines.

Specification:

Item

Specification

Input voltage

230VAC±10%, 50/60Hz, 3Amps

Output power of X ray tube

Max continuous output power 350W80kV/4.3mA or 43kV/8.0mA

Output voltage

Rated output voltageContinuously adjustable voltage range 30kV—80kV

Output voltage ripple±0.5%peak to peak

Output voltage accuracy±2% of voltage setting value

line regulation±0.1%

load regulation±0.1%

Tube current

Rated tube currentContinuously adjustable current range 0.5mA-8.0mA

Tube current accuracy±0.2mA of current setting value

line regulation±0.1%

load regulation±0.1%

Rise time of output voltage

The kV rise time is <0.6 Sec at maximum power

The kV rise time is <0.1 Sec at low voltage ( < 40kV )

Filament power supply:

input voltage24VDC

filament voltage2.0 to 5.3Vac

filament3.0 to 3.8 Amps RMS

preheating time3sec

Tube feature

Tube typefixed anodeberyllium window tube tungsten target

focus0.8mm

inherent filtration0.8mm Be

radiation angle45ºcone beam

target angle25°

Cooling

transformer oilexternal circulationexternal forced air cooling

Working temperatures

-10℃—40℃

Storing temperature

-20℃—60℃

System temperature protection

60℃±3℃ of oil temperature

Humidness

98%Non-condensation

Weight

Oil tank24.8kg

Control Unit3.8kg

Installation direction

Installation in any direction

Radiation angle

cone angle of ray tapered beam 45°

X-ray leakage

Less than 0.5mR/hr at 5cm from the surface of the HVC80804.

JB1/AC~(AC Input Power Connector)

Pin

Signal

Parameter

1

L

live wire

2

N

Neutral line

3

G

PE

1.jpg

JB2/COM (DMR-9S interface definitions)

Pin

Signal

Parameter

1.4.6.7.8.9

N/C

No connect

2

TXD

Data transmit

3

RXD

Data receive

5

GND

Signal gnd

2.png

JB3/Interlock,(DMR-9P interface definitions)

Pin

Signal

Parameter

3/4/5/6/7/8/9

N/C

No connect

1

Interlock Out


2

Interlock In


3.png

Short connect pin1 and pin2 make X ray source normal operation. Typical connection

4.png

Led indicator:

ID

Color

Meaning

XrayOn

Yellow

indicate X ray on

ARC

Red

Arcing in oil tank

OT

Red

Over temperature

EP_Err

Red

Tube voltage error

IP_Err

Red

Tube current error

Power

Green

Power on

Tank size

6.png5.png